Bench Validation of a Handcrafted Prototype Catheter for Intra-gastric Pressure Monitoring

Fecha de publicación: Fecha Ahead of Print:

Autores Icesi

  • Alberto Federico Garcia Marin

    Autor

  • Julian Chica Yanten

    Autor

Participantes ajenos a Icesi

  • Sanchez, AI
  • Aldana, JL
  • Mejia, JH
  • Burbano, D
  • Salazar, C

Resumen

Background: Abdominal compartment syndrome is a sustained intra-abdominal pressure (IAP) >20 mm Hg associated with new organ dysfunction. In order to prevent its development and related complications, IAP monitoring should be performed in patients with risk factors. Although techniques for its monitoring have been developed, they are of high cost and not always available in low- and lower–middle-income countries. Therefore, we aim to develop and validate in a bench model a handcrafted catheter to be used as an alternative method to measure the intra-gastric pressure (IGP) as a surrogate of the IAP. Methods: We used an acrylic water container as a model of the abdomen and four handcrafted catheters made of a 16 Fr Levin tube with a globe finger tied with silk in the distal end, inflated with 1 cm of air. They were placed on the bottom of the container where the water pressure was directly measured as a gold standard. The agreement between the two measures was assessed with the Bland–Altman method. Results: We performed 120 simultaneous measures. The mean pressure difference was 0.218 (95% CI 0.074 to 0.363). Conclusions: The handcrafted prototype catheter and the direct measure were highly correlated. The new catheter is a reliable and reproducible tool for pressure monitoring. However, before it can be used in the clinical setting for IAP monitoring, validation in human models in a real clinical setting needs to be performed. © 2020, Société Internationale de Chirurgie.

Datos de la publicación

ISSN/ISSNe:
0364-2313, 1432-2323

WORLD JOURNAL OF SURGERY  Springer Verlag

Tipo:
Article
Páginas:
1706-1711
PubMed:
32016541
Enlace a otro recurso:
www.scopus.com
Factor de Impacto:
1.115 SCImago ℠
Cuartil:
Q1 SCImago ℠

Citas Recibidas en Web of Science: 1

Citas Recibidas en Scopus: 2

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • abdomen; catheter; devices; human; intraabdominal hypertension; manometry; pathophysiology; physiologic monitoring; pressure; Abdomen; Catheters; Humans; Intra-Abdominal Hypertension; Manometry; Monitoring, Physiologic; Pressure