Impact of Using a Novel Gastric Feeding Tube Adaptor on Patient's Comfort and Air Leaks During Non-invasive Mechanical Ventilation
Fecha de publicación:
Fecha Ahead of Print:
Autores Icesi
Participantes ajenos a Icesi
- Quintero, OI
- Sanchez, AI
- Chavarro, PA
- Casas, IC
Unidades de Investigación
Resumen
Introduction: The presence of oral or naso-enteral probes during non-invasive mechanical ventilation (NIMV) increases the risk of leakage and patient discomfort. The objective of this study was to evaluate the effectiveness of a novel tube adapter for NIMV (TA-NIMV) in relation to leakage and comfort level. Methods: A non-randomized quasi-experimental design was performed in an adult intensive care unit of a highly complex hospital, in which patients were their own controls. We included adult patients who required NIV with oronasal mask and who simultaneously had oral or naso-enteric tubes. The interventions were as follows: every participant received two therapies, one with the TA-NIMV and one conventional therapy of NIMV (CT-NIMV). Comfort could be evaluated in 99 patients with a Glasgow Coma Scale of 15. The outcomes of interest was the average percentage of air leak and patient comfort during each intervention. Results: 196 patients were included in the study during a 16-month period. The mean air leak percentage was 9.2% [standard deviation (SD), 7.7] during TA-NIMV and 32.5% (SD, 12.5) during CT-NIMV (p < 0.001). 84.9% reported being comfortable or very comfortable during TA-VMNI. 66.7% Uncomfortable or Very uncomfortable during CT-NIMV (p < 0.001). Conclusion: Higher comfort levels and lower air leakage volume percentages were achieved using the TA-NIMV than those achieved by CT-NIMV. © 2019 The Author(s)
Copyright © 2020. Publicado por Elsevier España, S.L.U.
Datos de la publicación
- ISSN/ISSNe:
- 0300-2896, 1579-2129
- Tipo:
- Article
- Páginas:
- 353-359
- PubMed:
- 31732357
- Enlace a otro recurso:
- www.scopus.com
- Factor de Impacto:
- 0.287 SCImago ℠
ARCHIVOS DE BRONCONEUMOLOGIA Elsevier Doyma
Citas Recibidas en Web of Science: 9
Citas Recibidas en Scopus: 17
Documentos
- No hay documentos
Filiaciones
Keywords
- Gastric feeding tube; Intensive care units; Leak; Non-invasive ventilation; Patient comfort
Cita
Quintero OI,Sanchez AI,Chavarro PA,Casas IC,Ospina GA. Impact of Using a Novel Gastric Feeding Tube Adaptor on Patient's Comfort and Air Leaks During Non-invasive Mechanical Ventilation. Arch. Bronconeumol. 2019. 56. (6):p. 353-359. IF:4,957. (1).
Portal de investigación